Sum the table and see whether it reaches 0.9 per cent. A 99.1 per cent headline leaves 0.9 per cent of the detected area unaccounted for, and a related-substances table exists to say what that 0.9 per cent is. If the listed peaks add to 0.54 per cent, the missing 0.36 is unassigned area — real signal the laboratory saw and did not name — and that is a finding, not an omission. Check three things in order: does the table report each impurity as area per cent on the same basis as the headline; is there a reporting threshold below which peaks were discarded; and does the largest single impurity have a limit against it. A table whose entries do not close on 0.9 per cent is not wrong, but it is incomplete in a way the headline conceals.
The single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
Compare purity within a single laboratory on the same method, never across laboratories.