More usefully, certified reference materials from national metrology institutes carry a purity statement and a statement of traceability, and both matter.
The response factor changes if the method changes — a different gradient time, a different temperature, a different column — so the standard must be run under the exact method conditions as the sample.
The pharmacopoeial standards specify system suitability requirements for quantitative methods that include replicate injection agreement and that agreement should be tighter than two per cent.
One qualification: using a standard from a reputable source and checking its purity on receipt is worth the effort.
The practical summary: use a standard of declared purity and verified traceability, and check that the certificate is current.
5This should probably be in the site help pages rather than buried in an answer. – RP_C18 3 months ago 4Good answer, but the confidence interval in the cited trial is wider than implied. – petra_hovland 2 months ago add a comment