Accepted answer
The single most useful thing to do is write the arithmetic on the vial label, because you will reconstruct it from memory at an inconvenient moment if you do not.
Dead space quantified: a fixed-needle insulin syringe holds roughly 3 to 5 µL in the hub and needle after the plunger bottoms out. A luer-lock syringe with a detachable needle holds 35 to 100 µL depending on the hub design. At 5 mg/mL that is 15 to 25 µg lost per draw on the insulin syringe and 175 to 500 µg on the luer-lock — which over ten draws is the difference between losing a rounding error and losing half a milligram.
Dead space by syringe type
| Configuration | Dead volume | Loss at 5 mg/mL | Over 20 draws |
|---|
| Fixed-needle insulin syringe | 3–5 µL | 15–25 µg | 0.3–0.5 mg |
| Low-dead-space, detachable | <2 µL | <10 µg | <0.2 mg |
| Standard luer-lock + 30G | 35–60 µL | 175–300 µg | 3.5–6 mg |
| Luer-lock + 21G drawing needle | 70–100 µL | 350–500 µg | 7–10 mg |
Concretely, do not use the same needle to pierce the stopper and to administer. The tip is blunted by the stopper, and the hub now contains a dose you are about to lose to dead space anyway.
Published data on syringe dead space quantifies low-dead-space designs as retaining under 2 µL against 35 µL or more for conventional detachable-needle syringes.
If in doubt, use more diluent and accept the shorter usable window.
4Thank you — the worked example is what makes this usable. – b_delacroix 7 months ago 3Related: the same reasoning applies to the counter-ion question. – tandem_gradient 6 months ago add a comment