The honest answer is that the achievable range of plausible purity figures for a given vial is wider than most people expect.
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
Put another way, mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
Compare purity within a single laboratory on the same method, never across laboratories.