Accepted answer
98.4 per cent is a statement about area, and the other 1.6 per cent is everything the detector saw and did not assign to your peak. Read it as 98.4 of every 100 units of peak area at whatever wavelength was used, not as 98.4 per cent of the mass in the vial. With the batch certificate attached you can at least see how the figure was produced, which is the difference between a measurement and a claim. What it still does not tell you is content: how many milligrams are actually there.
The method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
It helps to be literal here: sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
Compare purity within a single laboratory on the same method, never across laboratories.